Description
A low-cost, research-grade data logger for continuous monitoring of water level and temperature at depths of up to 13, 30, or 100 feet in a wide range of underwater environments; for use in freshwater or saltwater.
A low-cost, research-grade data logger for continuous monitoring of water level and temperature at depths of up to 13, 30, or 100 feet in a wide range of underwater environments; for use in freshwater or saltwater.
Pressure (Absolute) and Water Level Measurements U20L-01 |
|
Operation Range | 0 to 207 kPa (0 to 30 psia); approximately 0 to 9 m (0 to 30 ft) of water depth at sea level, or 0 to 12 m (0 to 40 ft) of water at 3,000 m (10,000 ft) of altitude |
Factory Calibrated Range |
69 to 207 kPa, 0° to 40°C
(10 to 30 psia, 32° to 104°F) |
Burst Pressure |
310 kPa or 18 m depth
(45 psia or 60 ft depth) |
Water Level Accuracy* |
Typical error: ±0.1% FS, 1.0 cm (0.03 ft) water
Maximum error: ±0.2% FS, 2.0 cm (0.06 ft) water
|
Raw Pressure Accuracy** | ±0.3% FS, 0.62 kPa (0.09 psi) maximum error |
Resolution |
<0.02 kPa water
(0.007 ft water) |
Pressure Response Time (90%)*** | <1 second at a stable temperature; measurement accuracy also depends on temperature response time |
Pressure (Absolute) and Water Level Measurements U20L-02 |
|
Operation Range | 0 to 400 kPa (0 to 58 psia); approximately 0 to 30.6 m (0 to 100 ft) of water depth at sea level, or 0 to 33.6 m (0 to 111 ft) of water at 3,000 m (10,000 ft) of altitude |
Factory Calibrated Range |
69 to 400 kPa (10 to 58 psia), 0° to 40°C (32° to 104°F)
|
Burst Pressure | 500 kPa (72.5 psia) or 40.8 m (134 ft) depth |
Water Level Accuracy* |
Typical error: ±0.1% FS, 3.0 cm (0.1 ft) water
Maximum error: ±0.2% FS, 6.0 cm (0.2 ft) water
|
Raw Pressure Accuracy**
|
±0.3% FS, 1.20 kPa (0.17 psi) maximum error |
Resolution | <0.04 kPa (0.006 psi), 0.41 cm (0.013 ft) water |
Pressure Response Time (90%)*** | <1 second at a stable temperature; measurement accuracy also depends on temperature response time |
Pressure (Absolute) and Water Level Measurements U20L-04 |
|
Operation Range | 0 to 145 kPa (0 to 21 psia); approximately 0 to 4 m (0 to 13 ft) of water depth at sea level, or 0 to 7 m (0 to 23 ft) of water at 3,000 m (10,000 ft) of altitude |
Factory Calibrated Range |
69 to 145 kPa (10 to 21 psia), 0° to 40°C (32° to 104°F)
|
Burst Pressure |
310 kPa (45 psia) or 18 m (60 ft) depth
|
Water Level Accuracy*
|
Typical error: ±0.1% FS, 0.4 cm (0.013 ft) water
Maximum error: ±0.2% FS, 0.8 cm (0.026 ft) water
|
Raw Pressure Accuracy**
|
±0.3% FS, 0.43 kPa (0.063 psi) maximum error |
Resolution |
<0.014 kPa (0.002 psi), 0.14 cm (0.005 ft) water |
Pressure Response Time (90%)*** | <1 second at a stable temperature; measurement accuracy also depends on temperature response time |
Temperature Measurements (All Models) |
|
Operation Range |
-20° to 50°C (-4° to 122°F) |
Accuracy | ±0.44°C from 0° to 50°C , see manual for plot (±0.79°F from 32° to 122°F , see manual for plot) |
Resolution |
0.10°C at 25°C , see manual for plot
(0.18°F at 77°F , see manual for plot) |
Response Time (90%) | 10 minutes in water (typical) |
Stability (Drift) |
0.1°C per year (0.18°F per year) |
Logger |
|
Real-time Clock | ±1 minute per month 0° to 50°C (32° to 122°F) |
Battery | 2/3 AA, 3.6 Volt lithium, factory-replaceable |
Battery Life (Typical Use) | 5 years with 1 minute or greater logging interval |
Memory (Non-volatile) | 64K bytes memory (approx. 21,700 pressure and temperature samples) |
Weight |
Approximately 154 g (5.43 oz) in air
Approximately 53.9 g (1.9 oz) in fresh water
|
Dimensions | 3.18 cm (1.25 inches) diameter, 15.24 cm (6.0 inches) length; mounting hole 6.3 mm (0.25 inches) diameter |
Wetted Materials | Polypropylene housing and lanyard; Viton and Buna-N O-rings; ceramic sensor in acetal end cap; stainless steel screws suitable for saltwater |
Logging Interval | Fixed-rate or multiple logging intervals, with up to 8 user-defined logging intervals and durations; logging intervals from 1 second to 18 hours. Refer to the HOBOware User’s Guide for details. |
Launch Modes | Immediate start and delayed start |
Offload Modes | Offload while logging; stop and offload |
Battery Indication | Battery voltage can be viewed in status screen and optionally logged in datafile. Low battery indication in datafile. |
Environmental Rating | IP68 |