Pressure and Water Level Measurements U20-001-01 and U20-001-01-Ti
|
Operation Range |
0 to 207 kPa (0 to 30 psia); approximately 0 to 9 m (0 to 30 ft) of water depth at sea level, or 0 to 12 m (0 to 40 ft) of water at 3,000 m (10,000 ft) of altitude |
Factory Calibrated Range |
69 to 207 kPa (10 to 30 psia), 0° to 40°C (32° to 104°F)
|
Burst Pressure |
310 kPa (45 psia) or 18 m (60 ft) depth
|
Water Level Accuracy* |
Typical error: ±0.05% FS, 0.5 cm (0.015 ft) water
Maximum error: ±0.1% FS, 1.0 cm (0.03 ft) water
|
Raw Pressure Accuracy** |
±0.3% FS, 0.62 kPa (0.09 psi) maximum error |
Resolution |
<0.02 kPa (0.003 psi), 0.21 cm (0.007 ft) water
|
Pressure Response Time (90%)*** |
<1 second; measurement accuracy also depends on temperature response time |
Temperature Measurements (All Models)
|
Operation Range |
-20° to 50°C (-4° to 122°F)
|
Accuracy |
±0.44°C from 0° to 50°C (±0.79°F from 32° to 122°F), see Plot A in manual
|
Resolution |
0.10°C at 25°C (0.18°F at 77°F), see Plot A in manual
|
Response Time (90%) |
5 minutes in water (typical)
|
Stability (Drift)
|
|
Logger
|
Real-time Clock |
± 1 minute per month 0° to 50°C (32° to 122°F) |
Barometric Pressure (MX2001-TOP)
|
Operation Range |
66 to 107 kPa
(9.57 to 15.52 psia)
|
Temperature Calibrated Range
|
|
Accuracy |
±0.2 kPa over full temperature range at fixed pressure; maximum error ±0.5% FS
(±0.029 psi over full temperature range at fixed pressure; maximum error ±0.5% FS)
|
Water Level Accuracy*
|
Typical error: ±0.075% FS, 0.3 cm water
(Typical error: ±0.075% FS, 0.01 ft water)
Maximum error: ±0.15% FS, 0.6 cm water
(Maximum error: ±0.15% FS, 0.02 ft water)
|
Raw Pressure Accuracy**
|
±0.3% FS, 0.43 kPa (0.063 psi) maximum error |
Resolution |
|
Response Time |
<1 second at stable temperature |
Stability (Drift) |
<0.01 kPa per year
(<0.0015 psi per year) |
Logger
|
Real-time Clock |
± 1 minute per month 0° to 50°C (32° to 122°F) |
Battery |
2/3 AA, 3.6 Volt lithium, factory-replaceable |
Battery Life (Typical Use) |
5 years with 1 minute or greater logging interval |
Memory (Non-volatile) |
64K bytes memory (approx. 21,700 pressure and temperature samples) |
Weight |
Stainless steel models: approximately 210 g (7.4 oz)
Titanium models: approximately 140 g (4.8 oz) |
Dimensions |
2.46 cm (0.97 inches) diameter, 15 cm (5.9 inches) length; mounting hole 6.3 mm (0.25 inches) diameter |
Wetted Materials |
316 stainless steel, Viton® o-rings, acetyl cap, ceramic sensor |
Logging Interval |
Fixed-rate or multiple logging intervals, with up to 8 user-defined logging intervals and durations; logging intervals from 1 second to 18 hours. Refer to the HOBOware software manual. |
Launch Modes |
Immediate start and delayed start |
Offload Modes |
Offload while logging; stop and offload |
Battery Indication |
Battery voltage can be viewed in status screen and optionally logged in datafile. Low battery indication in datafile. |
Environmental Rating |
IP68 |
|
The CE Marking identifies this product as complying with all relevant directives in the European Union (EU). |