Description
Bluetooth-enabled logger
The industry’s first water level data logger designed for convenient wireless setup and download from mobile devices or Windows laptop via Bluetooth technology.
Distributor Data Logger & Solusi Sistem Integrator Berbagai Aplikasi Industri
Bluetooth-enabled logger
The industry’s first water level data logger designed for convenient wireless setup and download from mobile devices or Windows laptop via Bluetooth technology.
Pressure (Absolute) and Water Level Measurements MX2001-01-SS-S and MX2001-01-Ti-S |
|
Operation Range | 0 to 207 kPa; approximately 0 to 9 m of water depth at sea level, or 0 to 12 m of water at 3,000 m of altitude (0 to 30 psia; approximately 0 to 30 ft of water depth at sea level, or 0 to 40 ft of water at 10,000 ft of altitude) |
Factory Calibrated Range |
69 to 850 kPa, 0° to 40°C
(10 to 123.3 psia, 32° to 104°F)
|
Burst Pressure |
1,200 kPa or 112 m depth
(174 psia or 368 ft depth)
|
Water Level Accuracy* |
Typical error: ±0.05% FS, 3.8 cm water
(Typical error: ±0.05% FS, 0.125 ft water)
Maximum error: ±0.1% FS, 7.6 cm water
(Maximum error: ±0.1% FS, 0.25 ft water)
|
Raw Pressure Accuracy** |
±0.3% FS, 2.55 kPa maximum error
(±0.3% FS, 0.37 psi maximum error) |
Resolution |
<0.02 kPa, 0.21 cm water
(<0.003 psi, 0.007 ft water)
|
Pressure Response Time (90%)*** | <1 second at a stable temperature |
Pressure (Absolute) and Water Level Measurements MX2001-04-SS-S and MX2001-04-Ti-S |
|
Operation Range |
0 to 145 kPa; approximately 0 to 4 m of water depth at sea level, or 0 to 7 m of water at 3,000 m of altitude
(0 to 21 psia; approximately 0 to 13 ft of water depth at sea level, or 0 to 23 ft of water at 10,000 ft of altitude)
|
Factory Calibrated Range |
69 to 145 kPa, 0° to 40°C
(10 to 21 psia, 32° to 104°F)
|
Burst Pressure |
310 kPa or 18 m depth
(45 psia or 60 ft depth)
|
Water Level Accuracy* |
Typical error: ±0.075% FS, 0.3 cm water
(Typical error: ±0.075% FS, 0.01 ft water)
Maximum error: ±0.15% FS, 0.6 cm water
(Maximum error: ±0.15% FS, 0.02 ft water)
|
Raw Pressure Accuracy**
|
±0.3% FS, 0.43 kPa maximum error
(±0.3% FS, 0.063 psi maximum error)
|
Resolution |
<0.014 kPa, 0.14 cm water
(<0.002 psi, 0.005 ft water)
|
Pressure Response Time (90%)*** | <1 second at a stable temperature |
Barometric Pressure (MX2001-TOP) |
|
Operation Range |
66 to 107 kPa
(9.57 to 15.52 psia)
|
Temperature Calibrated Range
|
20 to 50°C
(-4 to 122°F)
|
Accuracy |
±0.2 kPa over full temperature range at fixed pressure; maximum error ±0.5% FS
(±0.029 psi over full temperature range at fixed pressure; maximum error ±0.5% FS)
|
Water Level Accuracy*
|
Typical error: ±0.075% FS, 0.3 cm water
(Typical error: ±0.075% FS, 0.01 ft water)
Maximum error: ±0.15% FS, 0.6 cm water
(Maximum error: ±0.15% FS, 0.02 ft water)
|
Raw Pressure Accuracy**
|
±0.3% FS, 0.43 kPa (0.063 psi) maximum error |
Resolution |
<0.01 kPa
(<0.0015 psi)
|
Response Time | <1 second at stable temperature |
Stability (Drift) |
<0.01 kPa per year
(<0.0015 psi per year) |
Temperature Measurements (All Sensor End Models MX2001-0x-SS-S and MX2001-0x-Ti-S) |
|
Operation Range |
-20° to 50°C
(-4° to 122°F)
|
Accuracy |
±0.44°C from 0° to 50°C, see manual for plot
(±0.79°F from 32° to 122°F, see manual for plot) |
Resolution |
0.1°C at 25°C, see manual for plot
(0.18°F at 77°F, see manual for plot)
|
Response Time (90%) | 5 minutes in water (typical) |
Stability (Drift) |
0.1°C per year
(0.18°F per year)
|
Logger |
|
Real-time Clock |
-20° to 50°C
(-4° to 122°F) |
Radio Power |
1 mW
(0 dBm) |
Transmission Range |
Approximately 30.5 m line-of-sight
(Approximately 100 ft line-of-sight) |
Wireless Data Standard | Bluetooth Low Energy (Bluetooth Smart®) |
Logging Rate |
1 second to 18 hours
|
Logging Modes |
Fixed interval, multiple intervals with up to 8 user-defined logging intervals and durations, or burst
|
Memory Modes
|
Wrap when full or stop when full
|
Start Modes
|
Immediate, date & time, or next interval
|
Stop Modes
|
When memory full, date & time, or after a set logging period
|
Time Accuracy
|
±1 minute per month 0° to 50°C
(±1 minute per month 32° to 122°F) |
Battery |
Two AA, 1.5 V alkaline batteries, user replaceable
|
Battery Life
|
1 year, typical with logging interval of 1 minute. Faster logging and/or statistics sampling intervals, entering burst logging mode, excessive readouts, checking of Full Status Details, and remaining connected with the app will impact battery life.
|
Memory | 256 KB memory (30,000 sets of measurements) |
Full Memory Download Time
|
Approximately 2 minutes; may take longer the further the device is from the top end of the logger |
Dimensions |
Top end (MX2001-TOP): 2.54 cm (1.0 inches) diameter, 28.9 cm (11.4 inches) length; mounting hole 7.6 mm (0.3 inches) diameter
Sensor end (MX2001-0x-SS-S and MX2001-0x-Ti-S: 2.54 cm (1.0 inches) diameter, 9.91 cm (3.9 inches) length
Note: The length of the water level logger cable (CABLE-DR-xxx) can vary -0% to +3% +10 cm (3.9 inches) from the length ordered. The logger adds 38.8 cm (15.3) inches to the length of the cable ordered.
|
Weight | Top end (MX2001-TOP): Approximately 136 g (4.78 oz) in air Stainless sensor end (MX2001-0x-SS-S): Approximately 141.4 g (4.99 oz) in air; approximately 93.4 g (3.3 oz) in fresh water Titanium sensor end (MX2001-0x-Ti-S): Approximately 80 g (2.83 oz) in air; approximately 37 g (1.3 oz) in fresh water |
Wetted Materials | Top end (MX2001-TOP): PVC housing, Polycarbonate end cap, Viton O-ring Stainless sensor end (MX2001-0x-SS-S): Stainless steel housing, Viton and Buna-N O-rings, ceramic sensor in stainless steel end cap Titanium sensor end (MX2001-0x-Ti-S): Acetal housing, Viton and Buna-N O-rings, ceramic sensor in Titanium end cap Cable (CABLE-DR-XXX): Polycarbonate strain relief (top end connection), PVC end cap (sensor connection), nylon collar nuts, Viton O-rings, polyurethane jacket |
Environmental Rating
|
|
Notices |
|
* | Water Level Accuracy: With accurate reference water level measurement, known water density, and a stable temperature environment. System Water Level Accuracy equals the sum of the Barometric Water Level Accuracy plus the selected sensor end Water Level Accuracy. |
** | Raw Pressure Accuracy: Absolute pressure sensor accuracy includes all sensor drift, temperature, and hysteresis-induced errors. |
*** | Changes in Temperature: Allow 20 minutes in water to achieve full temperature compensation of the pressure sensor. There can be up to 0.5% of additional error due to rapid temperature changes. Measurement accuracy also depends on temperature response time. |
Reviews